Analytical Chemistry 4 分で読了 945 語

UV-Vis 分光法とランベルト-ベール則

光吸収で濃度を測定する

Principles of UV-Vis Spectroscopy

Ultraviolet-visible (UV-Vis) spectroscopy is one of the most widely used analytical techniques in chemistry. It measures how much light a sample absorbs at different wavelengths in the ultraviolet (200–400 nm) and visible (400–700 nm) regions of the electromagnetic spectrum. Because many chemical compounds absorb light at characteristic wavelengths, UV-Vis spectroscopy serves both as a qualitative identification tool and — more powerfully — as a quantitative concentration measurement method.

When a photon of light encounters a molecule, it can be absorbed if its energy matches the energy gap between the molecule's ground electronic state and an excited state. This excitation most commonly involves: - π → π transitions in compounds with C=C double bonds, aromatic rings, and conjugated systems - n → π** transitions in compounds containing lone pairs adjacent to double bonds (C=O, N=O groups)

The wavelength at which a compound shows maximum absorbance is called λ_max (lambda max) and is a characteristic property of the compound's electronic structure.

The Beer-Lambert Law

The mathematical relationship between absorbance and concentration is described by the Beer-Lambert Law (also called Beer's Law):

A = ε × c × l

Where: - A = absorbance (dimensionless; unitless logarithmic ratio) - ε (epsilon) = molar absorptivity (L mol⁻¹ cm⁻¹) — a constant characteristic of the compound at a given wavelength - c = concentration (mol/L or M) - l = path length of the cuvette (cm; typically 1.00 cm)

Absorbance is defined as:

A = log₁₀(I₀ / I)

Where I₀ is the incident light intensity and I is the transmitted light intensity. Transmittance (T = I/I₀) is related by A = −log₁₀(T).

Key Implications

  • Beer's Law predicts a linear relationship between absorbance and concentration — the foundation of quantitative analysis.
  • A plot of absorbance vs. concentration (a calibration curve) should be a straight line passing through the origin with slope ε × l.
  • The relationship is valid at low to moderate concentrations. At high concentrations, intermolecular interactions cause deviations from linearity.

Instrumentation

A UV-Vis spectrophotometer contains: 1. Light source: A deuterium lamp for UV (200–400 nm) and a tungsten-halogen lamp for visible (400–800 nm) 2. Monochromator: A diffraction grating or prism that selects a specific narrow wavelength band 3. Sample compartment: Holds a cuvette (optical cell) containing the sample 4. Detector: A photodiode or photomultiplier tube converts transmitted light intensity to an electrical signal 5. Data system: Computes absorbance and generates spectra

Cuvettes are made of quartz (for UV measurements, since glass absorbs below 300 nm) or optical-grade glass/plastic (for visible-only work). They must be handled carefully — fingerprints and scratches on optical faces cause serious errors.

Diode array spectrophotometers collect the entire spectrum simultaneously rather than scanning wavelength-by-wavelength, enabling rapid kinetic measurements.

Quantitative Analysis Procedure

  1. Select λ_max: Run a full spectrum of the analyte and identify the wavelength of maximum absorbance. Using λ_max maximizes sensitivity and minimizes error from small wavelength variations.

  2. Prepare standards: Make a series of solutions of known concentration spanning the expected range of the unknown.

  3. Measure absorbances: Record A for each standard at λ_max, blanking the instrument with a solvent reference.

  4. Construct a calibration curve: Plot A vs. concentration. Fit a linear regression (y = mx + b). A good calibration curve has R² ≥ 0.999.

  5. Measure the unknown: Read A for the unknown and calculate its concentration from the regression equation.

Example: A student dissolves an iron complex (ε = 11,100 L mol⁻¹ cm⁻¹ at 510 nm) in a 1.00 cm cuvette and reads A = 0.385.

c = A / (ε × l) = 0.385 / (11,100 × 1.00) = 3.47 × 10⁻⁵ mol/L

Limitations and Deviations from Beer's Law

Beer's Law is an approximation with known limitations:

  • Concentration deviations: At high concentrations (typically > 0.01 mol/L), electrostatic interactions alter ε, causing negative deviations (curve bends below the line).
  • Polychromatic radiation: If the monochromator's bandwidth is too wide and ε varies significantly across the band, apparent deviations occur.
  • Stray light: Unwanted radiation reaching the detector causes apparent absorbances that are too low at high true absorbances.
  • Chemical equilibria: If the absorbing species participates in a pH-dependent equilibrium (e.g., indicators), the measured spectrum changes with conditions.

For accurate work, absorbances between 0.1 and 1.5 are ideal. Very low absorbances (< 0.05) suffer from poor signal-to-noise; very high absorbances (> 2.0) approach the detector's detection limit.

Qualitative Applications

Beyond quantitation, UV-Vis spectra provide structural information:

  • Aromatic compounds show characteristic π → π* bands. Benzene absorbs at ~254 nm; naphthalene at ~312 nm; anthracene at ~375 nm.
  • Conjugated dienes and polyenes: increasing conjugation shifts λ_max to longer wavelengths (bathochromic shift). β-carotene (11 conjugated double bonds) absorbs visible light near 450 nm — hence its orange color.
  • Protein quantification: Proteins absorb at 280 nm due to tryptophan and tyrosine residues. DNA absorbs at 260 nm (aromatic bases). The A₂₈₀/A₂₆₀ ratio helps assess purity.

Real-World Applications

UV-Vis spectroscopy permeates science and industry:

  • Clinical diagnostics: Glucose, bilirubin, hemoglobin, and enzyme activity in blood are measured photometrically in clinical analyzers.
  • Environmental analysis: Nitrate in drinking water (at 220 nm), phosphate as molybdenum blue complex, and ozone in air (UV absorption at 254 nm).
  • Pharmaceutical quality control: Assaying drug concentrations in tablets dissolved in solvent.
  • Kinetics: Monitoring reaction progress by following absorbance change over time, allowing calculation of rate constants.
  • Remote sensing: Satellite instruments use UV-Vis spectroscopy to monitor atmospheric ozone, NO₂, and SO₂ concentrations globally.

The combination of simplicity, speed, low cost, and the power of the Beer-Lambert Law makes UV-Vis spectroscopy an enduring cornerstone of analytical chemistry.